GOSSEN Foto- und Lichtmesstechnik GmbH 2 This document describes the measurement procedure for MAVOSPEC in combination with MAVOSPHERE 205,
GOSSEN Foto- und Lichtmesstechnik GmbH 3 2 The calibration of the self absorption Before further steps, the support for the sphere contr
GOSSEN Foto- und Lichtmesstechnik GmbH 4 In order to enable sphere controller port click the box "Sphere controller settings" and then
GOSSEN Foto- und Lichtmesstechnik GmbH 5 When the proper adapter is mounted to the MAVOSPEC, the Menu "Sphere Correction" will be disp
GOSSEN Foto- und Lichtmesstechnik GmbH 6 Measurement screen.
GOSSEN Foto- und Lichtmesstechnik GmbH 7 In the second step install the DUT. The measured light source (DUT) has to be switched off. Only the
GOSSEN Foto- und Lichtmesstechnik GmbH 8 The new correction file will be available on the Sphere correction list in Quick Config panel.
GOSSEN Foto- und Lichtmesstechnik GmbH 9 3 The actual measurement of light source Now the device is ready for the actual measuring of the light